Abstract

Single photon emission computed tomography (SPECT) is an important biomedical imaging modality. However, since gamma cameras are expensive and bulky, truncated projection data are either preferred or unavoidable. Inspired by the recent results on interior tomography in the x-ray CT field, here we present the interior SPECT approach for exact and stable reconstruction of a region of interest (ROI) from uniformly attenuated local projection data, aided by prior knowledge of a sub-region in the ROI. First, by analytic continuation we prove that interior SPECT is both exact and stable, and by singular value decomposition (SVD) we establish the stability of interior SPECT. Then, given the constant attenuation coefficient and object boundary, our interior SPECT reconstruction is achieved by inverting a generalized truncated Hilbert transform using the SVD technique. Preliminary numerical simulation data demonstrate that our work has practical utilities. The theoretical generalization of our work to the variable attenuation case is underway, and the same numerical approach can be applied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.