Abstract

The homogeneity test of glass plates in a Fizeau interferometer requires the measurement of the glass sample in reflected as well as in transmitted light. For the measurement in transmitted light, the sample has to be inserted into the ray path of a Fizeau or Twyman-Green interferometer, which leads to a nested cavity setup. To separate the interference signals from the different cavities, we illuminate a Fizeau interferometer with an adaptive frequency comb. In this way, rigid glass plates can be measured, and linear variations in the homogeneity can also be detected. The adaptive frequency comb is provided by a variable Fabry-Perot filter under broadband illumination from a superluminescence diode. Compared to approaches using a two-beam interferometer as a filter for the broadband light source, the visibility of the fringe system is considerably higher.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.