Abstract

The theory of certain errors involved in interferometric frequency calibration is developed. The origins of these errors are (1) the effective slit width of the monochromator and (2) the height of the slit. For uncoated, low reflecting interferometer plates a complete theory is given. For higher reflecting plates the effect of the slit height has been calculated in detail for one definite choice of reflection coefficient. The results of this calculation are thought to be representative also for other cases.The main effect both of the slit width and the slit height is a blurring of the interference pattern. But moreover the maxima are displaced on account of the slit height effect. This may cause serious errors in calibrations. Curves are presented, giving the “fractional order” of the maxima for different cases.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.