Abstract

Recent advances in laser driven accelerators boosted the development of high dose-rate fast pulsed systems all over the world. The stray radiation comprises primarily high-energy photons, resulting in extremely high dose rates within pico-/femto-second pulses. Dose measurements in such conditions have to be evaluated to validate methods. To tackle this challenge the EUropean RAdiation DOSimetry Group (EURADOS) started a program of dosimeters intercomparison, with a progressive approach, starting by a first evaluation in fields with μs pulse duration.The first comparison took place at the Lausanne University Hospital Center with an electron LINAC in Sept. 2017 involving 7 European institutes. Several passive and active dosimeters were tested with a tunable air kerma per pulse of the order of MGy/h. All instruments, except electrets, did not show any dose rate dependence, thus being selected as possible candidates for further studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.