Abstract

Interactive MASI involves controlling the raster of a STEM or SEM probe to areas predefined byan integration mask which is formed by image processing, drawing or selecting regions manually. EELS, x-ray, or other spectra are then acquired while the probe is scanning over the areas defined by the integration mask. The technique has several advantages: (1) Low-dose spectra can be acquired by averaging the dose over a great many similar features. (2) MASI can eliminate the risks of spatial under- or over-sampling of multiple, complicated, and irregularly shaped objects. (3) MASI is an extremely rapid and convenient way to record spectra for routine analysis. The technique is performed as follows:Acquire reference imageOptionally blank beam for beam-sensitive specimensUse image processor to select integration mask from reference imageCalculate scanning path for probeUnblank probe (if blanked)Correct for specimen drift since reference image acquisition

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.