To assess inter- and intraobserver variability in wireless resonance frequency analysis (RFA) stability measurements of palatal implants and to evaluate the influence of age, sex, time after implant insertion, and measurement direction on variability. Three observers conducted wireless RFA stability measurements of palatal implants from 16 patients. Measurements were taken in anteroposterior and laterolateral directions and were repeated after 1 hour. Data showed a small interobserver variation (1.93) with intraobserver variation (1.77) as its largest component. Time after implantation showed a strong influence (P = .027) on the interobserver variation. The tested wireless RFA device can be considered a reliable instrument for measuring the stability of palatal implants. Multiple readings and observers could further improve measurement reliability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.