A system of instrumentation for continuous measurement of oxygen consumption (VO2), carbon dioxide production (VCO2) and minute ventilation (V1) in human newborns on assisted ventilation is described. VO2 and VCO2 are measured by open-circuit indirect calorimetry utilizing a Servomex OA 184 differential paramagnetic oxygen analyzer and a Beckman LB 2 infrared CO2 analyzer. Minute volume is measured with a body plethysmograph. Bench performance is described, limitations of the system are defined, and sample clinical data are presented. The instrumentation can measure safely, accurately, and continuously physiologic variables in sick infants on assisted ventilation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.