Abstract

The relative radiation intensity ( R i ) defined as fluorescent radiation intensity of analyte in specimen to fluorescent radiation intensity of pure element or compound, e.g., oxide is used in calculation in both fundamental parameter methods and in theoretical influence coefficient algorithms. Accuracy of calculated R i is determined by uncertainties of atomic parameters, spectrometer geometry and also by X-ray tube spectral distribution. This paper presents the differences between R i calculated using experimental and theoretical X-ray tube spectra evaluated by three different algorithms proposed by Pella et al., Ebel, and Finkelshtein–Pavlova. The calculations are performed for the most common targets, i.e., Cr, Mo, Rh and W. In this study, R i is calculated for V, Cr, Mn, Fe, Co, Ni, Cu and Mo in steels as an example. The differences between R i calculated using different X-ray tube spectrum algorithms are presented when pure element standard, multielement standard similar to the analyzed material and one pure element standard for all analytes is used in X-ray fluorescence analysis. The differences between R i for intermediate-thickness samples (and also for thin films) and for X-ray tube, which ran for many hours, are also evaluated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.