Abstract
We have found that to the detection sensitivity of Total reflection X-ray fluorescent spectrometry (TXRF), the total volume of trace particles generated by vapor phase treatment (VPT) must be increased and metal atoms need to be included in the particles. The detection sensitivity for Cu is enhanced by assisting Cu ionization in the liquid drops condensed form the vapor. We consider that since incident and reflected X-rays resonate 30nm from the surface, the total reflection intensity of metals included in the particles is enhanced.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.