Abstract

To extend pH stability, protective metal oxide layers, such as titanium oxide, that are more stable in alkaline medium, can be chemically bonded to the chromatographic silica surface prior to reaction with silanes. In the present work, the influence of the titanium oxide content on the chromatographic performance was investigated by synthesizing a C 18 phase onto a doubly-titanized silica support and comparing its chromatographic performance with a C 18 phase on singly-titanized silica. The Engelhardt and Tanaka test mixtures were used for chromatographic characterizations using short HPLC columns. The column lifetimes of these titanized phases were also compared by performing accelerated aging tests at 50 °C using aggressive phosphate mobile phases at pH 10.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.