Abstract

Results from a Monte Carlo investigation are presented for the application of the dual photopeak window (DPW) scatter correction method to phantom geometries of varying shapes, sizes, and densities as well as to several collimators. A Monte Carlo program was used to simulate a Tc-99m point source at locations of 0.5 cm from the phantom surface, at the center, 0.5 cm from the bottom, and 4/5 the distance lateral from the center for each case of variation. DPW was then used to correct for scatter using a regression relation determined for a 30.5 cm/spl times/23.0 cm water-filled elliptical tub phantom using an ultrahigh resolution collimator. The application of DPW in all cases accounted for an approximate twelve-fold reduction in scatter fraction with an excellent agreement between the true scatter and scatter estimate distributions both at the tails and under the peak of the point spread functions. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.