Abstract
Objective To explore the influence of dose rate reference control level on the design of accelerator shielding. Methods According to the standards of GBZ 126-2011, GBZ/T 201.1-2007 and GBZ/T 201.2-2011, two different shielding calculation methods, based on the dose rate reference control level and week dose control level, were used to calculate and compare the thickness of accelerator room shielding. Results Under the same condition of maximum weekly workload, the obtained results were different when the calculated dose rate reference control value was larger than dose rate reference control level (2.5 μSv/h). The maximum difference of shielding thickness reached 64 cm. Meanwhile, considering dose rate reference control level, the different accelerator rates could lead to different radiation shielding thickness. Conclusions The dose rate at reference point must be first calculated before calculating shielding thickness. The calculation should be made on the premise that dose rate reference control level is met. Key words: Accelerator; Shielding; Dose rate reference control level
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.