Abstract

The pre- and post-irradiation scan strategy for optical-CT gel readout often turns out to be corrupted by angular mismatch between these two scans. In this study, we used computational simulations to investigate the influence of angular mismatch. Two phantoms are constructed: one cylindrical phantom with synthetic impurities and one elliptical phantom. The reconstructed results of angular mismatched pre- and post-data show that the dual-scan method is very sensitive to repositioning error, and positive-negative pair errors can be easily identified around impurities and phantom edges. From the simulation results, we believe that the angular mismatch should be less than 0.1 degree.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.