In order to investigate the influence of abutment material on the stress of implant-supported all-ceramic single crown, a 3D finite element model of implant-supported mandibular first premolar was computed by COSMOS/M 2.85 software. Alumina, zirconia, and titanium were used as abutment materials respectively. Vertical 600 N and horizontal 225 N load was applied on the occlusal surface. The results show that the stress distribution of implant-supported single crown was similar for different abutment materials. Maximum stresses within the crown were higher when titanium abutment was used. Maximum stress of titanium abutment was lower than that of ceramic abutment. Within the screw and fixture, maximum stresses had no difference under vertical loading but higher as titanium abutment was used under horizontal loading. There was no difference of maximum stress within the bone when different abutment materials were used. The present findings indicate that the abutment material had no influence on the stress distribution of implant-supported allceramic single crown but maximum stress when the titanium abutment was lower than that of ceramic abutment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.