Abstract

Laboratory of National Measurement Standard for Electricity and Time (NMS ET Lab) has validated the measurement results pre and post the LCR meter recalibration process on inductance parameters. The validation process was carried out at nominal values of 100 μH, 1 mH, 10 mH, 100 mH and 1 H in accordance to the reference standards. The direct measurement method was used to measure the inductance parameter of LCR meter with standard inductor. Validation is carried out by comparing the measurement results pre and post the LCR meter recalibration process at SCL - Hong Kong. Based on the E n number analysis, the values between -0.21 to 0.44 were obtained for 5 measurement points, therefore the calibration system and method used by NMS ET Lab are valid and have measurement results that are equivalent to other countries.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.