Abstract

In totally implantable ventricular assist device systems, measuring flow rate of the pump is necessary to ensure proper operation of the pump in response to the recipient's condition or pump malfunction. To avoid problems associated with the use of flow probes, several methods for estimating flow rate of a rotary blood pump used as a ventricular assist device have been studied. In the present study, we have performed a chronic animal experiment with two NEDO PI gyro pumps as the biventricular assist device for 63 days to evaluate our estimation method by comparing the estimated flow rate with the measured one every 2 days. Up to 15 days after identification of the parameters, our estimations were accurate. Errors increased during postoperation days 20 to 30. Meanwhile, their correlation coefficient r was higher than 0.9 in all the acquired data, and estimated flow rate could simulate the profile of the measured one.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.