A new incoherent Thomson scattering system measures the evolution of electron velocity distribution functions perpendicular and parallel to the ambient magnetic field during kinking of a single flux rope and merging of two flux ropes through magnetic reconnection. The Thomson scattering system provides sub-millimeter spatial resolution, sufficient to diagnose the several millimeters sized magnetic reconnection electron diffusion region in the PHAse Space MAppgin experiment. Due to the relatively modest plasma density ∼1019m-3 and electron temperature ∼1eV, stray light suppression is critical for these measurements. Two volume Bragg gratings are used in series as a notch filter with a spectral bandwidth <0.1nm in the collection branch. A CCD with a Gen III intensifier with peak quantum efficiency >47% is used as the detector in a 1.3m spectrometer. Preliminary results of gun plasma electron temperature will be reported and compared with measurements obtained from a triple Langmuir probe.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.