Abstract

Experimental findings are presented of an in vivo comparison between a Sentron catheter and another tip transducer manometer: a Millar microtip catheter. Both catheters have been used simultaneously in the left ventricle of dogs. Pressure variations were elicited by drug infusion. Pressure values and derivatives obtained from both systems were compared. A cross correlation between episodes of the two pressures was computed. Results from this study showed good correlation between left ventricular systolic pressure measured with both manometers (R = 0.992, p < 0.0001), end-diastolic pressure (R = 0.809, p < 0.0001) and between first derivatives: positive derivative (R = 0.993, p < 0.0001) and negative (R = 0.634, p < 0.0001). The mean cross correlation between both pressure signals was 0.61 +/- 0.04. In the frequency domain no statistical difference was found between the location of the maxima of the peaks. It is concluded that a Sentron manometer can be a valid alternative, at a reasonable price, to a cheaper, though less accurate fluid filled catheter and a more expensive 'golden standard' microtip catheter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.