Abstract

In this study, the corrosion and surface properties of non-toxic and non-allergenic β-type Ti–29Nb–13Ta–4.6Zr alloys (TNTZ) coated with TiN, ZrN, and CrN layers using the PVD technique were investigated under in-vitro conditions. The coatings' hardness, adhesion resistance, and antibacterial properties were also studied, with a focus on their potential usage as biomaterials. The corrosion tests were conducted using potentiodynamic polarization, the hardness tests were conducted using the Knoop method, and the adhesion tests were conducted using pull-off tests. The wetting properties of the coatings were examined through contact angle and 3D morphological analysis. The coatings were characterized using SEM-EDS and XRD. The results showed that the adhesion resistance of all coatings was significantly higher than the expected adhesion strength for biomaterial coatings, while their corrosion resistance was higher compared to the uncoated alloy. Besides, the study demonstrates the potential of TiN, ZrN, and CrN coatings on β-type TNTZ for use as a biomaterial with desirable surface and antibacterial properties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.