Abstract

This study investigated the strain surround implants and the pressure distribution of the alveolar ridge related to implant overdentures with stress-breaking ball (SBB) attachments. Simulation models of edentulous mandibles were embedded with 1 and 2 implants to simulate a single implant-retained overdenture (S-IOD) and a two-implant-retained overdenture (T-IOD), respectively. The implants were connected to the denture base with the following conditions: complete separation between the denture base and implant with cover screws as a conventional complete denture (CD), 3 types of settlements with an SBB attachment between them, and a conventional ball attachment (BALL). Strain surround the implant, pressure at 5 different soft tissue areas, and displacement of the denture base were measured with loading up to 50 N. In S-IOD and T-IOD, the strain of each SBB attachment was less than the strain shown by the BALL. The pressure at each region of the SBB attachment was less than the pressure shown with the CD and BALL. SBB attachments prevented strain surround the implants in loading and provided optimal stress distribution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.