A novel method for collecting and processing coincidence data from a Compton Suppressed Low Energy Photon Spectrometer (LEPS) is presented, greatly simplifying the current setup and extending the suppression abilities of the system. Offline analysis is used, eliminating the need to discard coincidence data when vetoing coincident events with fast-timing electronics. Additional coincident events are identified that are usually missed, and which represent interactions in the active NaI(Tl) shield prior to an interaction in the LEPS detector. By suppressing these events, the Compton Suppression factor was improved by 144% for the 661.66keV decay line in a 137Cs source. The geometry used for this particular Compton suppression system is highly sensitive to these effects, however similar event profiles are expected in all coincidence systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.