Abstract

Acid etch bonding of orthodontic attachments to the labial and lingual surfaces of teeth is generally adequate for clinical service. Failures do occur, and these are related to technique problems in acid etching or resin manipulation or to bracket base design. The widely used foil mesh base can present problems in retention because of design defects or corrosion. Improvement in mechanical retention of resin to the attachment base would result in significant improvements in bonding and fewer clinical failures. One of our approaches to improved retention is to fuse metallic or ceramic particles onto the bracket base so as to achieve a particulate or porous layer into which resin can penetrate. The results of bond-strength tests showed that a 100 percent increase in bond strength could be obtained with sintered porous metal-coated brackets. Further development is proceeding.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.