Abstract
Studies were made of the anlaytical conditions required for indirect photometric ion chromatography using ultraviolet photometric detection (UV method) for the determination of serum cations following a previously developed serum pre-treatment. The sensitivities of the conductivity detection (CD) and UV methods and the amounts of serum cations determined by both methods were compared. Attempts to improve the sensitivity of the conventional UV method are reported. It was found that the mobile phase previously reported by Small and Miller showed no quantitative response when more than 4 m M copper (II) sulphate pentahydrate was used. As a result, there was no significant difference in the amounts of serum cations shown by the CD and UV methods. However, by adding 0.5–5 m M cobalt(II) sulphate heptahydrate, nickel(II) sulphate hexahydrate, zinc(II) sulphate heptahydrate or cobalt(II) diammonium sulphate hexahydrate to 0.5–1.5 m M copper(II) sulphate pentahydrate, higher sensitivity and a quantitative response were attained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.