Abstract

An improved UV spectrophotometric method has been developed and validated for precise, efficient and selective determination of dexamethasone in tablets and capsules. The quantitative analyses were carried out using ethanol/water (2:1 v/v) as background electrolyte and UV detection was carried out at 240 nm. The calibration curve was linear over a concentration range from 4.0 to 40.0 µg mL -1 . The average recovery was 97.60 ± 1.06% for tablets and 96.64 ± 0.87% for capsules. The limit of detection and limit of quantification were 0.63 and 1.90 µg mL -1 , respectively. The method was validated according to ICH guidelines and the acceptance criteria for accuracy, precision, linearity, specificity and system suitability were met. The results obtained with proposed method confirm improved performance over other methods found in the literature . DOI: http://dx.doi.org/10.17807/orbital.v7i1.630

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.