Abstract
An improved liquid feed system for the Berglund-Liu vibrating orifice monodisperse aerosol generator (VOMAG) is described. The feed system employs a positive-displacement pump and pulse dampener to provide an accurate solution feed rate to the VOMAG. Advantages over the syringe pump feed system are reduced orifice clogging, elimination of leaks, and large solution reservoir for increased run times. Solution feed rate to the VOMAG using this feed system was reproducible to +/-2%. The system was low cost and required only a simple calibration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.