Abstract

We recently designed, built and commissioned a new pinhole / filter assembly for the equatorial hard x-ray imager (eHXI) at the National Ignition Facility (NIF). In this paper we describe the design and metrology of the new diagnostic as well as the spectral and spatial response of the hard x-ray detector. The new eHXI assembly has improved the photon collection efficiency along with spectral and spatial resolution by making use of 1D imaging channels and various hard x-ray filters. In addition we added a Ross pair filter set for Au K-alpha emission (67-69 keV). The new eHXI design will improve our understanding of the sourcing of hot electrons, generated in laser-plasma-instabilities, along the vertical hohlraum axis. This information is an important input for simulating and eventually limiting the DT fuel preheat in ICF implosions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.