Abstract

We clinically evaluated, retrospectively, our improved fluorescence polarization assay for fetal lung maturity. The procedure requires 0.5 mL of amniotic fluid and a standard clinical laboratory fluorescence polarimeter (TDx Analyzer, Abbott Laboratories). We measured the L/S ratios for 93 freshly collected amniotic fluids, uncontaminated with blood or meconium, collected within three days of delivery. The fluids were stored frozen for eight to 32 months, then thawed and assayed for net fluorescence polarization. Fourteen of the infants developed respiratory distress syndrome; five, transient tachypnea of the newborn; and 74, no respiratory distress. The polarization assay and lecithin/sphingomyelin ratio had equivalent receiver operating characteristic curves, indicating no difference in their clinical performance. Although a prospective study with fresh amniotic fluid specimens will be necessary to establish a definitive reference range, the present study shows that this assay can be used to rapidly predict fetal lung maturity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.