Abstract
Introduction: Automatic Capture Verification (ACV) technology enables advancement in automated and remote patient management. The reliability of evoked response (ER) sensing via reduced output coupling capacitor technology has been clinically proven. This technology was applied to the LV and used in this study to evaluate the effect of interventricular pacing timing or LV Offset (LVO) during biventricular (BiV) pacing for a new ACV algorithm. Methods: Sinus rhythm patients (pts) indicated for CRT-D implantation were enrolled for intraoperative evaluation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.