Abstract
We have realized an improved resistance calibration setup based on a cryogenic current comparator (CCC). The new comparator with 18 windings and a total of 4647 turns suits well for all the necessary comparisons of a 100-Omega standard resistor with the quantum Hall effect (QHE) and within the range of standard resistance values from 1 Omega to 1 MOmega. The new state-of-the-art setup is equipped with a dc superconducting quantum interference device (SQUID) and a digital double current source, allowing a reduction in the ramping time by at least one order of magnitude and significantly improving the dynamic stability of the system. The latter also benefits from the internal wideband feedback that we present in this paper.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.