Abstract

Left ventricular assist devices (LVADs) are carefully designed, but the significance of the implantation configuration and interaction with the vasculature is complex and not fully determined. The present study employs computational fluid dynamics to investigate the importance of applying a realistic LVAD profile when evaluating assisted aortic flow fields and subsequently compares a number of potential anastomosis locations in a patient-specific aortic geometry. The outflow profile of the Berlin Heart INCOR® device was provided by Berlin Heart GmbH (Berlin, Germany) and the cannula was attached at a number of locations on the aorta. Simulations were conducted to compare a flat profile against the real LVAD profile. The results illustrate the importance of applying an LVAD profile. It not only affects the magnitude and distribution of oscillatory shear index, but also the distribution of flow to the great arteries. The ascending aorta was identified as the optimal location for the anastomosis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.