We report on results of high precision straightness measurements of a graduated scale and a comparison of two different error separation methods. One approach is based on the Traceable Multi-Sensor method. For this purpose, a sensor element consisting of three heterodyne interferometers was integrated at PTB's length reference comparator to measure the straightness deviation of structures on scales. For a scale with a grating of 322 mm length a standard uncertainty below 1.5nm was achieved using this method. A comparison with another method based on reversal technique was realized and revealed a difference below 2.5nm (PV) between both results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.