Abstract
We are developing, at the Huazhong University of Science and Technology, a small-animal positron emission tomograph (PET) scanner that has two new features. Firstly, it uses detector modules that are based on a new digital readout electronics previously investigated and reported by us. Secondly, it can arrange the detector modules into various scanner geometries for different imaging needs. In this paper, we will describe the design and implementation of the detector module and report its performance properties. The detector module has a total detection sensitive area of 5.3 × 5.3 cm2, an overall energy resolution of 14.1% FWHM at 511 keV, and a block-level coincidence timing resolution of 1.5 ns FWHM. In addition, our preliminary imaging study shows that our small-animal PET scanner can resolve 1.0 mm-diameter rods separated by 2.0 mm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.