Abstract

Displacement of Impella 5.0 secondary to patient movement or transportation is a well-known complication. Typically, repositioning of an Impella across the aortic valve is attempted over a guidewire. We present the first case, to our knowledge, of repositioning a dislodged Impella 5.0 without a guidewire under transesophageal echocardiography (TEE) guidance, by inducing rapid ventricular pacing to cross the aortic valve. A 70-year-old man with low left ventricular ejection fraction underwent off-pump coronary artery bypass grafting. On 2nd postoperative day a low cardiac output state developed with increasing lactate levels and consequently the patient was taken to the cardiac catheterization laboratory for insertion of an Impella 5.0. Suddenly the Impella system failed with a rapid hemodynamic deterioration and it was successfully bedside repositioned inducing rapid ventricular pacing. In case of accidental Impella dislodgement and fast deterioration of the patient's hemodynamic status, rapid pacing may be an option to "open" the aortic valve thus aiding quick replacement of Impella 5.0 through the aortic valve into the left ventricle under TEE guidance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.