Abstract

To compare mean impedance levels, neural response telemetry (NRT), and auditory perception after initial and explant-reimplant pediatric cochlear implants. Retrospective case review. Tertiary referral hospital and cochlear implant program. Children 0 to 16 years inclusive who have undergone explant-reimplant of their cochlear implant. Impedance levels, NRT, and speech perception performance. Impedance, NRT, and auditory perception at switch on, 3 months, 12 months, 3 years, and 5 years after initial cochlear implant and reimplantation. The explant-reimplant group receiving Cochlear contour array had significantly (p < 0.001) raised impedance at switch on, 3 months, 12 months, and 3 years, compared with their initial implant. The explant-reimplant group receiving Cochlear straight array had marginally significant (p = 0.045) raised impedance at switch on, 3 months, 12 months, and 3 and 5 years. Infection was associated with greater increases in impedance in the reimplant Contour group. NRT was increased in the explant-reimplant group but not significantly (p = 0.06). Auditory perception returned to preexplant levels within 6 months in 61% of children. Impedance is higher after explant-reimplant and remains increased for years after explant-reimplant with Cochlear contour and to a lesser degree the straight array device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.