Abstract
Helical computed tomography (HCT) has become the preferred protocol in routine clinical applications. The advantages of HCT include its capability of scanning a complete anatomical volume in a single breath-hold, and the capability of generating images at arbitrary locations. However, studies have indicated various HCT-related image artifacts in 3D applications. In this paper, we perform a detailed analysis on the cause of one of the artifacts in the 3D shaded surface display (SSD): the grooving artifact. We developed an analytical model to show that the artifact is induced by the angular-dependent edge response of the HCT reconstruction. The accuracy of the model is demonstrated by both computer simulations and phantom experiments. Based on the model, we present several correction algorithms to combat the grooving artifacts. Computer simulations and phantom experiments have shown the effectiveness and thoroughness of the correction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.