Abstract

In this paper, the impact of the electrode structure on radio-frequency (RF) induced heating near an active implantable medical device implanted lead under magnetic resonance imaging exposure is investigated. The specific absorption rate (SAR) distributions and the temperature rises of the leads with different electrodes were assessed. It is shown that increasing the size of the lead electrode reduces the SAR distribution and temperature rise near the electrode. Our results indicate that a larger electrode size will reduce the magnitudes of the lead transfer function, subsequently reducing the heating effect near the electrode. Both numerical simulations and experimental measurements were performed to verify the effectiveness of a large electrode in mitigating the RF-induced heating.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.