Abstract

ObjectivesTo assess the potential influence of the ratio between the storage tube surface area and the volume of cerebrospinal fluid (CSF) (surface/volume) on the quantifications of four Alzheimer's disease (AD) biomarkers on the Lumipulse G600II automated platform. MethodsCSF samples of 10 consecutive patients were stored in 2 ml polypropylene tubes containing four different CSF volumes: 1.5 ml, 1 ml, 0.5 ml and 0.25 ml. Concentration of CSF Aβ1-42, Aβ1–40, t-Tau and p-Tau were measured in all aliquots using the LUMIPULSE G600II automated platform from Fujirebio. ResultsLevels of CSF Aβ1-42 and Aβ1-40 were lower in samples stored with lower volumes (higher surface/volume ratios). This decrease was partly compensated by using the ratio Aβ1-42/Aβ1-40. Quantification of t-Tau and p-Tau were not influenced by this pre-analytical condition. ConclusionThe surface/volume ratio can potentially influence the results of amyloid AD biomarkers. It appears essential to take into account the surface/volume ratio of the storage tubes when quantifying CSF biomarkers in clinical routine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.