Abstract
Reduction of priming volumes of the cardiopulmonary bypass (CPB) circuit in neonatal cardiac surgery to decrease haemodilution and blood transfusion requirements can be achieved with the use of neonatal low prime oxygenators and smaller diameter tubing. We have further reduced our prime volume with the use of a custom-designed arm allowing for remote positioning of a double-headed roller pump. This arm enables the double pump to be placed alongside the main heart-lung machine close to the operating table, and to position the pump inlet and outlet tubing immediately at the reservoir outlet and oxygenator inlet, respectively, therefore reducing tubing lengths. Priming volumes of four cases using this configuration were compared to four cases using our standard neonatal bypass setup. Results showed a 29% decrease in priming volume and a 58% reduction in blood utilization during CPB. This reduction in priming volume is clinically significant as it lowers the ratio of priming volume to patient blood volume and reduces homologous blood requirements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.