The rendering of low contrast values by various mammographic imaging systems was investigated by a semi-objective procedure. Test exposures of a line raster with variable line widths were evaluated visually. Significantly better results were achieved by a xero-radiographic method. This was better than non-screen mammographic film in terms of object size, lower sensitivity to scatter and better resolution. The use of a tungsten rotating anode, with about 0.5 mm. Al filter, in xeromammography provided a satisfactory compromise between image quality and radiation dose in the rendering of low contrast studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.