Abstract

The increased utilization of continuous glucose monitors (CGM) and smart insulin pens (SIP) among people with type 2 diabetes generates significant health data. This study explored possible patterns in long term CGM and SIP data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.