Abstract

Incorrect registration during computer assisted total knee arthroplasty (CA-TKA) leads to malposition of implants. Our aim was to evaluate the tolerable error in anatomic landmark registration. We incorrectly registered the femoral epicondyles, femoral and tibial centers, as well as the malleoli and documented the change in angulation or rotation. We found that the distal femoral epicondyles were the most difficult anatomic landmarks to register. The other bony landmarks were more forgiving. Identification of the distal femoral epicondyles has a high inter-observer and intra-observer variability. Our observation that there is less than 2mm of safe zone in the anterior or posterior direction during registration of the medial and lateral epicondyles may explain the inability of CA-TKA to improve upon the outcomes of conventional TKA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.