Abstract

Liver function reserve (LFR) plays an extensive and important role in patients with liver disease. Indocyanine green (ICG) clearance test is the standard diagnostic approach for LFR evaluation which was performed by spectrophotometry or pulse dye densitometry (PDD). Spectrophotometry is the gold standard, it's invasive and not real-time. PDD is non-invasive, but accuracy of PDD is controversial. Taken spectrophotometry as the reference standard, this study investigated the accuracy of photoacoustic imaging (PAI) method for LFR assessment and compared to PDD in healthy volunteers. The results demonstrated a strong correlation between PAI method and spectrophotometry (r = 0.9649, p < 0.0001). No significant difference was shown in ICG clearance between PAI and spectrophotometry method (rate constant k1 vs. k2, 0.001158 +−0.00042 vs. 0.001491 +- 0.00045, p = 0.0727; half-life t1 vs. t2, 601.2 s vs. 474.4 s, p = 0.1450). These results indicated that PAI may be valuable as a noninvasive, accurate diagnostic tool for LFR assessment in human.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.