Abstract

A fast, sensitive and specific hydrophilic interaction chromatography combined with tandem mass spectrometry (HILIC-MS/MS) method was developed for the determination of tobramycin in human plasma. With sisomycin as internal standard, the analysis was carried out on a hilic column (150mm×2.1mm, 3.5μm) using a mobile phase consisting of acetonitrile: 5mM ammonium acetate and 0.1% formic acid (60:40, v/v). The detection was performed by tandem spectrometry via electrospray ionization (ESI). Linear calibration curves were obtained in the concentration range of 10.51-1051ng/mL for tobramycin, with a lower limit of quantification of 10.51ng/mL. The intra- and inter-day precision (RSD) values were below 15% and accuracy (RE) was 1.3-5.7% at all QC levels. The method was applicable to the clinical study of the pharmacokinetics of tobramycin in healthy volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.