Abstract
Continuous passive motion (CPM) and continuous active motion (CAM) physiotherapy devices are commonly used to promote rehabilitation of damaged synovial joints. CAM devices use some form of mechanical resistance to patient muscle activity. This paper presents a combined CAM/CPM device which functions conventionally in the CPM mode, but in the CAM mode uses the motor to exert a constant force against patient activity, simulating the effect of a spring load. In the CAM mode, the motor is embedded in a feedback loop with two sensors, one a force sensor, the other an angular position sensor. The PC provides full recording and graphic display of the patients force and, angular position of the device and motor force. The simulated-spring-load system operation is smooth and natural. The presented design will be of use for diagnosis, therapy and research.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.