Abstract

The purpose of this study was to evaluate the accuracy of clinical measurement of cervical dilation with a position-tracking system during vaginal examination. This prospective study that was conducted in Poissy, France, Brooklyn, NY, and Haifa, Israel, included 333 measurements that were performed in 188 women with term singleton vertex uncomplicated pregnancies during the active stage of labor. Ninety measurements with clinical diagnosis of full dilation were excluded from analysis. Measurements were performed with a sensor attached to the midwife's index fingertip and a position-tracking system that was based on a low magnetic field. Evaluations were done when cervical examinations were clinically indicated. Results were similar in all centers. Mean error was 10.2 +/- 8.4 mm and ranged from 7.5 +/- 7.3 mm, when cervical dilation was > 8 cm, to 12.5 +/- 8.7 mm when cervical dilation was between 6.1 and 8 cm. This first evaluation of cervical assessment accuracy during vaginal examination with a position-tracking system shows limited precision.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.