Abstract

The aim of this study is to quantitatively evaluate the influence of the duration of ambient light exposure on the image quality of digital radiographs obtained with contemporary phosphor plate (PSP)-based systems. Radiographs of an aluminum step-wedge were obtained using VistaScan and Express systems at five X-ray exposure times: 0.10, 0.20, 0.32, 0.40, and 0.50 s; the resulting dose-area products were, respectively, 21.93, 43.87, 70.19, 87.75, and 109.6 mGycm2. Before PSP read-out, half of the sensitive surface of the PSP plates was exposed to ambient light for 5, 10, 30, 60, and 90 s. The effect of light exposure on brightness, contrast, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and image saturation was compared using ANOVA with the Tukey test (α = 0.05). Ambient light exposure increased brightness and contrast and reduced CNR and SNR in PSP-based radiographs of contemporary digital systems. At the longest X-ray exposure times, ambient light exposure reduced the dark saturation (mean gray values ≤ 1) observed in Express. In conclusion, the negative effects of ambient light observed on the image quality of PSP-based radiographs are directly proportional to the duration of exposure. Clinicians should be aware of such harmful effects when handling and scanning PSP plates in bright environments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.