Abstract

In this study, implantable polyethylene/fluorouracil (PE/FU) rods were manufactured by hot extrusion under different processing conditions. SEM–EDS analyses revealed the effect of temperature on the morphology of the samples. Furthermore, small particles of fluorouracil were observed on the surface and in the PE matrix. Both the FTIR and NIR spectra of the PE/FU rods confirmed the presence of fluorouracil. The PE/FU rods presented lower values of flexural modulus and fatigue resistance than pure PE rods; this was probably due to imperfections and defects introduced into the PE matrix by the fluorouracil particles. The initial amount of FU released by the extruded PE/FU rods (around 35 mg/g) is desirable since it provides a high initial concentration of the drug locally to kill cancer cells following implantation. The subsequent slow and controlled release of the drug (12–45 days) provides suitable levels of the chemotherapeutic agent at the tumor site to improve the anticancer treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.