Abstract

Heat and moisture exchange filters (HMEFs) have been a controversial yet widely used adjunct to breathing circuit. Here we discuss a case scenario where clogging of HMEF with moisture presented as elevated peak airway pressures. Apart from stressing the importance of monitoring airway pressure, we discuss about the judicious use of HMEFs, problems associated with their use and measures to overcome the same. Being a ubiquitous adjunct, a thorough knowledge of its functioning and constant vigilance during its use is of paramount importance

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.