Abstract

We investigated the performance of the new p24 antigen detection kit (VIDAS HIV p24) with the conventional antigen kit (HIV-1 Ag monoclonal; Abbott). The new kit is an enzyme-linked fluorescent immunoassay (ELFA) and all of the assay steps are performed automatically by the VIDAS instrument within 100 minutes. With the seven HIV-1 seroconversion panels, three seroconversions were detected on an average of 6.8 days earlier with ELFA than the conventional EIA kit. ELFA showed negative results for all of the 11 false positive samples by the combined (p24, anti-HIV) detection kit (VIDAS HIV DUO). The results obtained suggest that ELFA are very useful for an earlier diagnosis of HIV infection and re-test for false positive samples by other HIV diagnosis kits.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.