Abstract
Spike and wave activity is detected by analogue circuitry which separately filters the spike and the slow wave components from the EEG. The filter outputs are converted to DC voltages, which are then compared with individually adjusted threshold voltages. A detector output pulse is produced when both the spike and the wave threshold levels are exceeded. A graph of the incidence of spike and wave activity per 15 min intervals over 24 h is plotted by a microcomputer and modified impact printer. The analysis is carried out at 60 times real time on EEG recordings obtained from unrestrained patients using a portable cassette-tape recorder.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Electroencephalography and Clinical Neurophysiology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.